Trends in patient dose and compression force for digital (DR) mammography systems over an eleven-year period

Phys Eng Sci Med. 2024 Mar;47(1):215-222. doi: 10.1007/s13246-023-01357-x. Epub 2023 Nov 29.

Abstract

This study evaluated trends in patient dose and compression force for screening digital (DR) mammography systems. The results of five audits (carried out in 2011, 2014, 2018, 2020 and 2022) were compared. For every audit, anonymised screening examinations from each system consisting of the standard craniocaudal (CC) and mediolateral oblique (MLO) views of both breasts were analysed. Exposure parameters were extracted from the Digital Imaging and Communications in Medicine (DICOM) header and the mean glandular dose (MGD) for each image was calculated. Trends in the distribution of MGD, compressed breast thickness, compression force and compression force per radiographer were investigated. The mean MGD per image (and mean compressed breast thickness) was 1.20 mGy (58 mm), 1.53 mGy (59 mm), 1.83 mGy (61 mm), 1.94 mGy (60 mm) and 2.11 mGy (61 mm) for 2011, 2014, 2018, 2020 and 2022 respectively. The mean (and standard deviation) compression force was 114 (32) N, 112 (29) N, 108 (27) N, 104 (24) N and 100 (23) N for 2011, 2014, 2018, 2020 and 2022 respectively. The mean MGD per image has increased over time but remains below internationally established Diagnostic Reference Levels (DRLs). This increase is primarily due to a change in the distribution of the different manufacturers and digital detector technologies, rather than an increase in the dose of the individual systems over time. The mean compression force has decreased over time in response to client feedback surveys. The standard deviation has also reduced, indicating more consistent application of force.

Keywords: Breast screening; Compression force; Digital mammography; Mean glandular dose; Patient dose audit.

MeSH terms

  • Breast* / diagnostic imaging
  • Diagnostic Reference Levels
  • Humans
  • Mammography*
  • Physical Phenomena
  • Radiation Dosage