Inter- and intraindividual variation in Doppler ultrasonic indirect blood pressure measurements in healthy cats

J Vet Intern Med. 1999 Jul-Aug;13(4):314-8. doi: 10.1892/0891-6640(1999)013<0314:iaivid>2.3.co;2.

Abstract

This study was undertaken to evaluate reference ranges for systolic blood pressure (SBP) in cats under conditions mimicking a clinical setting. SBP was measured in 50 healthy adult cats of various ages (range, 1.5-16 years) and body weights (range, 2.2-6.1 kg) by Doppler ultrasonic sphygmomanometry. A cuff width of 2.5 cm was used, placed on the left antebrachium, and this represented a mean cuff width of 35% limb circumference (range, 31-42%). The mean (+/-SD) SBP in the 50 cats was 162 +/- 19 mm Hg (range 124-210), with only 1 cat having a SBP > or = 200 mm Hg. No significant difference (P > .05) in SBP was found between male and female cats, and no significant correlation was found between SBP and age (r(s) = 0.075) or body weight (r(s) = 0.007). Further studies in some of these cats indicated that allowing a period of 10 minutes for acclimatization to the environment where SBP was recorded resulted in a significant decrease in SBP from 176 +/- 17 to 157 +/- 21 mm Hg (n = 7) and that use of a 3.3-cm-width cuff resulted in a significant decrease in measured SBP from 168 +/- 13 to 164 +/- 13 mm Hg (n = 10). Reproducibility of SBP measurements was evaluated in 7 cats by assessing SBP 7 times at intervals of > or = 24 hours over a 10-day period. These 7 cats had a low intraindividual coefficient of variation of SBP measurements (CV < or = 7.9%) although 2 of the 7 cats had SBP values > 200 mm Hg on at least 1 occasion.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Blood Pressure
  • Blood Pressure Determination / veterinary
  • Cats
  • Female
  • Hypertension / diagnosis
  • Hypertension / veterinary*
  • Male
  • Reference Values
  • Reproducibility of Results
  • Sphygmomanometers
  • Ultrasonography, Doppler / veterinary*