Rapid cardiopulmonary support for children with complex congenital heart disease

Ann Thorac Surg. 2000 Sep;70(3):742-9; discussion 749-50. doi: 10.1016/s0003-4975(00)01562-9.

Abstract

Background: Extracorporeal membrane oxygenation has limitations in children with congenital heart disease (prolonged setup times, increased postoperative blood loss, and difficulty during transport). We developed a miniaturized cardiopulmonary support circuit to address these limitations.

Patients and methods: The cardiopulmonary support system includes a preassembled, completely heparin-coated circuit, a BP-50 Bio-Medicus centrifugal pump, a Minimax plus membrane oxygenator, a Bio-Medicus flow probe, and a Bio-trend hematocrit/oxygen saturation monitor. Short tubing length permits a 250-mL bloodless prime in less than 5 minutes. From 1995 to 1997, 23 children with congenital heart disease were supported with this technique.

Results: Overall survival to discharge was 48% (11 of 23 patients). Survival to discharge was 80% (4 of 5) in the preoperative support group, 20% (1 of 5) in the postoperative failure to wean from cardiopulmonary bypass group, 44% (4 of 9) in the group placed on support postoperatively after transfer to the intensive care unit, and 50% (2 of 4 patients) in the nonoperative group. Neonatal cardiopulmonary support survival to discharge was 46% (6 of 13 patients).

Conclusions: This pediatric cardiopulmonary support system is safe and effective. Advantages over conventional extracorporeal membrane oxygenation include rapid setup time, decreased postoperative blood loss, and simplified transport.

MeSH terms

  • Blood Transfusion
  • Cardiopulmonary Bypass / instrumentation*
  • Cardiopulmonary Bypass / methods
  • Cardiopulmonary Bypass / mortality
  • Equipment Design
  • Evaluation Studies as Topic
  • Heart Defects, Congenital / surgery*
  • Humans
  • Infant
  • Infant, Newborn
  • Transportation of Patients
  • Treatment Outcome