Automated responsiveness test (ART) predicts loss of consciousness and adverse physiologic responses during propofol conscious sedation

Anesthesiology. 2001 Apr;94(4):585-92. doi: 10.1097/00000542-200104000-00010.

Abstract

Background: The authors evaluated a device designed to provide conscious sedation with propofol (propofol-air), or propofol combined with 50% nitrous oxide (N2O; propofol-N2O). An element of this device is the automated responsiveness test (ART), a method for confirming that patients remain conscious. The authors tested the hypotheses that the ART predicts loss of consciousness and that failure to respond to the ART precedes sedation-induced respiratory or hemodynamic toxicity.

Methods: The protocol consisted of sequential 15-min cycles in 20 volunteers. After a 15-min control period, propofol was infused to an initial target effect-site concentration of 0.0 microg/ml with N2O or 1.5 microg/ml with air. Subsequently, the propofol target effect-site concentration was increased by a designated increment (0.25 and 0.5 microg/ml) and the process repeated. This sequence was continued until loss of consciousness, as defined by an Observer's Assessment of Alertness/Sedation (OAA/S) score of 10/20 or less, or until an adverse physiologic event was detected.

Results: The OAA/S score at which only 50% of the volunteers were able to respond to the ART (P50) during propofol-N2O was 11.1 of 20 (95% confidence interval [CI]: 10.6-11.8); the analogous P50 was 11.8 of 20 (95% CI: 11.4-12.3) with propofol-air. Failure to respond to the ART occurred at a plasma propofol concentration of 0.7 +/- 0.6 microg/ml with propofol-N2O and 1.6 +/- 0.6 microg/ml with propofol-air, whereas loss of consciousness occurred at 1.2 +/- 0.8 microg/ml and 1.9 +/- 0.7 microg/ml, respectively. There were no false-normal ART responses.

Conclusion: The ART can guide individual titration of propofol because failure to respond to responsiveness testing precedes loss of consciousness and is not susceptible to false-normal responses. The use of N2O with propofol for conscious sedation decreases the predictive accuracy of the ART.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Adult
  • Conscious Sedation*
  • Female
  • Hemodynamics / drug effects
  • Humans
  • Hypnotics and Sedatives / pharmacology*
  • Male
  • Probability
  • Propofol / adverse effects
  • Propofol / blood
  • Propofol / pharmacology*
  • Respiration / drug effects
  • Unconsciousness*

Substances

  • Hypnotics and Sedatives
  • Propofol