Platelet function point-of-care tests in post-bypass cardiac surgery: are they relevant?

Br J Anaesth. 2002 Nov;89(5):715-21.

Abstract

Background: Platelet dysfunction is an important cause of excessive bleeding after cardiac surgery. We assessed two platelet function point-of-care tests: the platelet function analyser (PFA-100) and the Hemostatus(TM) in patients with and without excessive bleeding after cardiac surgery with cardiopulmonary bypass.

Methods: Mediastinal chest tube drainage (MCTD) was measured for the first 6 h in the intensive care unit (ICU). Haematology and coagulation tests were done on arrival in the ICU, and when excessive bleeding occurred (MCTD >1 ml kg(-1) h(-1)) or after 3 h.

Results: Eighteen patients bled excessively and 27 had normal MCTD. Hemostatus measurements were prolonged in those with excessive bleeding compared with the normal group. The times for PFA-100 adenosine diphosphate (ADP) and epinephrine were 91 vs 71 s (P=0.004) and 155 vs 114 s (P=0.02) in the bleeding and normal group s, respectively. None of the Hemostatus or PFA-100 values correlated with total MCTD. Depending on the agonist used, maximum aggregation was 33-81% and 52-86% in bleeding and normal groups, respectively. Only poor correlations were found between PFA-100 epinephrine and maximum aggregation in response to ADP (r=-0.52, P=0.03) or to collagen (r=-0.48, P=0.04).

Conclusion: Patients bleeding excessively in the ICU had abnormal measurements in point-of-care tests without a dramatic decrease in aggregation. Except for patients with increased risk of postbypass bleeding, point-of-care tests are not useful for routine use after cardiac surgery.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Blood Platelets / physiology*
  • Cardiopulmonary Bypass*
  • Drainage
  • Female
  • Humans
  • Male
  • Middle Aged
  • Platelet Aggregation
  • Platelet Count
  • Point-of-Care Systems*
  • Postoperative Hemorrhage / etiology
  • Postoperative Hemorrhage / physiopathology*