Initial experience with a multiplane transoesophageal echo-transducer: assessment of diagnostic potential

Eur Heart J. 1992 Sep;13(9):1201-6. doi: 10.1093/oxfordjournals.eurheartj.a060338.

Abstract

Background: the prototype of a transoesophageal echocardiographic transducer with a rotatable cross-sectional scanning plane underwent initial evaluation.

Methods: the 5 MHz, phased array, 64 element transducer is incorporated into a 16 by 11 by 40 mm echoscope tip. The instrument also has pulsed wave and colour flow Doppler capabilities. Exterior controls allow continuous mechanical rotation of the scanning plane from 0 degree, corresponding to the conventional transverse plane, through 180 degrees, thereby encompassing all possible planes.

Results: 103 patients underwent examination without complications; two additional patients were excluded because of difficulty in swallowing the probe. Advantages include precise alignment of aortic valve long- and short-axis views, long-axis views of the ascending aorta (mean visualized length: 6 cm), and full scanning of the entire circumference of the mitral valve and the left ventricle. Separation of paravalvular and transvalvular leakage in prosthetic valves is distinctly improved.

Conclusion: multiplanar transoesophageal imaging is feasible and increases the diagnostic yield, especially in mitral and aortic pathology and in the assessment of left ventricular wall motion. Three-dimensional reconstruction is an attractive potential application.

MeSH terms

  • Adult
  • Aged
  • Algorithms
  • Computer Graphics
  • Echocardiography / instrumentation*
  • Equipment Design
  • Esophagoscopes
  • Female
  • Heart Diseases / diagnostic imaging*
  • Heart Valve Diseases / diagnostic imaging
  • Humans
  • Image Processing, Computer-Assisted / instrumentation*
  • Male
  • Middle Aged
  • Transducers*