Accuracy of MSCT coronary angiography with 64-slice technology: first experience

Eur Heart J. 2005 Aug;26(15):1482-7. doi: 10.1093/eurheartj/ehi261. Epub 2005 Apr 19.

Abstract

Aims: The aim of our study was to investigate the accuracy of 64-slice computed tomography (CT) for assessing haemodynamically significant stenoses of coronary arteries.

Methods and results: CT angiography was performed in 67 patients (50 male, 17 female; mean age 60.1+/-10.5 years) with suspected coronary artery disease and compared with invasive coronary angiography. All vessels > or =1.5 mm were considered for the assessment of significant coronary artery stenosis (diameter reduction >50%). Forty-seven patients were identified as having significant coronary stenoses on invasive angiography with 18% (176/1005) affected segments. None of the coronary segments needed to be excluded from analysis. CT correctly identified all 20 patients having no significant stenosis on invasive angiography. Overall sensitivity for classifying stenoses was 94%, specificity was 97%, positive predictive value was 87%, and negative predictive value was 99%.

Conclusion: Sixty-four-slice CT provides a high diagnostic accuracy in assessing coronary artery stenoses.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Calcinosis / diagnostic imaging
  • Coronary Angiography / methods
  • Coronary Angiography / standards*
  • Coronary Stenosis / diagnostic imaging*
  • Coronary Stenosis / physiopathology
  • Female
  • Heart Rate / physiology
  • Humans
  • Male
  • Middle Aged
  • Predictive Value of Tests
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / methods
  • Tomography, X-Ray Computed / standards*