Endoscopic removal of malfunctioning biliary self-expandable metallic stents

Gastrointest Endosc. 2005 Dec;62(6):903-10. doi: 10.1016/j.gie.2005.08.051.

Abstract

Background: Endoscopic removal of malfunctioning self-expandable metallic biliary stents (SEMS) is difficult and not well described. The aim of this study is to review the indications, the techniques, and the results of SEMS removal in a cohort of patients with malfunctioning stents.

Methods: All patients who underwent an attempt at endoscopic removal of biliary SEMS over a 5-year period were retrospectively identified. The main indications for SEMS removal were the following: distal migration of the stent or impaction to the duodenum, impaction into the bile-duct wall, tissue ingrowth, and inappropriate length of the stent causing occlusion of intrahepatic ducts. SEMS were removed by using foreign-body forceps or polypectomy snares.

Results: Endoscopic removal of 39 SEMS (13 uncovered and 26 covered) was attempted in 29 patients (17 men; mean age, 66 years). SEMS extraction was attempted after a mean of 7.5 months (8.75 months standard deviation) post-SEMS insertion. Removal was successful in 20 patients (68.9%) and in 29 SEMS (74.3%). Covered SEMS were effectively removed more frequently than uncovered ones: 24 of 26 (92.3%) and 5 of 13 (38.4%), respectively (p < 0.05). No major complications were recorded. Multivariate analysis showed that the time interval between insertion and removal, SEMS length, stent-mesh design (zigzag vs. interlaced), and indication for removal were not predictive of success at stent removal.

Conclusions: Endoscopic removal of biliary SEMS is feasible and safe in more than 70% of cases. Because only 38% of uncovered SEMS were removable, the presence of a stent covering is the only factor predictive of successful stent extraction. The presence of diffuse and severe ingrowth was the main feature limiting SEMS removal.

MeSH terms

  • Aged
  • Biliary Tract*
  • Cholestasis / etiology
  • Device Removal*
  • Endoscopy, Digestive System*
  • Equipment Failure
  • Female
  • Foreign-Body Migration / complications
  • Foreign-Body Migration / therapy
  • Humans
  • Male
  • Stents / adverse effects*