Mechanical evaluation of a new minimally invasive device for stabilization of proximal humeral fractures in elderly patients: a cadaver study

Acta Orthop. 2007 Jun;78(3):430-5. doi: 10.1080/17453670710014031.

Abstract

Background: Treatment of proximal humerus fractures in elderly patients is challenging because of reduced bone quality. We determined the in vitro characteristics of a new implant developed to target the remaining bone stock, and compared it with an implant in clinical use.

Methods: Following osteotomy, left and right humeral pairs from cadavers were treated with either the Button-Fix or the Humerusblock fixation system. Implant stiffness was determined for three clinically relevant cases of load: axial compression, torsion, and varus bending. In addition, a cyclic varus-bending test was performed.

Results: We found higher stiffness values for the humeri treated with the ButtonFix system--with almost a doubling of the compression, torsion, and bending stiffness values. Under dynamic loading, the ButtonFix system had superior stiffness and less K-wire migration compared to the Humerusblock system.

Interpretation: When compared to the Humerusblock design, the ButtonFix system showed superior biomechanical properties, both static and dynamic. It offers a minimally invasive alternative for the treatment of proximal humerus fractures.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Biomechanical Phenomena
  • Cadaver
  • Female
  • Fracture Fixation, Internal / instrumentation*
  • Fracture Fixation, Internal / methods
  • Humans
  • Male
  • Minimally Invasive Surgical Procedures / instrumentation*
  • Shoulder Fractures / surgery*
  • Weight-Bearing