Closed-loop control in fused MR-TRUS image-guided prostate biopsy

Med Image Comput Comput Assist Interv. 2007;10(Pt 1):128-35. doi: 10.1007/978-3-540-75757-3_16.

Abstract

Multi-modality fusion imaging for targeted prostate biopsy is difficult because of prostate motion during the biopsy procedure. A closed-loop control mechanism is proposed to improve the efficacy and safety of the biopsy procedure, which uses real-time ultrasound and spatial tracking as feedback to adjust the registration between a preoperative 3D image (e.g. MRI) and real-time ultrasound images. The spatial tracking data is used to initialize the image-based registration between intraoperative ultrasound images and a preoperative ultrasound volume. The preoperative ultrasound volume is obtained using a 2D sweep and manually registered to the MRI dataset before the biopsy procedure. The accuracy of the system is 2.3 +/- 0.9 mm in phantom studies. The results of twelve patient studies show that prostate motion can be effectively compensated using closed-loop control.

Publication types

  • Evaluation Study

MeSH terms

  • Biopsy, Needle / instrumentation*
  • Biopsy, Needle / methods
  • Equipment Design
  • Equipment Failure Analysis
  • Feedback
  • Humans
  • Image Interpretation, Computer-Assisted / instrumentation
  • Image Interpretation, Computer-Assisted / methods*
  • Imaging, Three-Dimensional / instrumentation
  • Imaging, Three-Dimensional / methods
  • Magnetic Resonance Imaging / instrumentation*
  • Magnetic Resonance Imaging / methods
  • Male
  • Prostatic Neoplasms / diagnostic imaging*
  • Prostatic Neoplasms / pathology*
  • Rectum / diagnostic imaging
  • Subtraction Technique / instrumentation
  • Surgery, Computer-Assisted / instrumentation*
  • Surgery, Computer-Assisted / methods
  • Ultrasonography / instrumentation*
  • Ultrasonography / methods
  • User-Computer Interface