Comparison of the Infiniti vision and the series 20,000 Legacy systems

Ophthalmologica. 2008;222(2):96-9. doi: 10.1159/000112625. Epub 2008 Feb 22.

Abstract

Purpose: To compare the efficiency of the Infiniti vision system and the Series 20,000 Legacy system phacoemulsification units during routine cataract extraction.

Methods: Thirty-nine eyes of 39 patients were randomized to have their cataract removed using either the Infiniti or the Legacy system, both using the Neosonix handpiece. System settings were standardized. Ultrasound time, amount of balanced salt solution (BSS) used intraoperatively, and postoperative visual acuity at postoperative days 1, 7 and 30 were evaluated.

Results: Preoperatively, best corrected visual acuity was significantly worse in the Infiniti group compared to the Legacy group (0.38 +/- 0.23 and 0.21 +/- 0.16, respectively; p = 0.012). The mean phacoemulsification time was 39.6 +/- 22.9 s (range 6.0-102.0) for the Legacy group and 18.3 +/-19.1 s (range 1.0-80.0) for the Infiniti group (p = 0.001). The mean amounts of intraoperative BSS used were 117 +/- 37.7 ml (range 70-195) in the Legacy group and 85.3 +/- 38.9 ml (range 40-200) in the Infiniti group (p = 0.005). No differences in postoperative visual acuity were found.

Conclusion: The ability to use higher flow rates and vacuum settings with the Infiniti vision system allowed for cataract removal with less phacoemulsification time than when using the Legacy system.

Publication types

  • Comparative Study
  • Evaluation Study
  • Randomized Controlled Trial
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Cataract Extraction / methods*
  • Cataract Extraction / standards
  • Humans
  • Intraocular Pressure
  • Middle Aged
  • Phacoemulsification / methods*
  • Phacoemulsification / standards
  • Postoperative Period
  • Time Factors
  • Treatment Outcome
  • Ultrasonic Therapy* / instrumentation
  • Ultrasonic Therapy* / standards
  • Visual Acuity