Flow estimation using ultrasound imaging (color M-mode) and computer postprocessing

J Cereb Blood Flow Metab. 1991 Sep;11(5):879-82. doi: 10.1038/jcbfm.1991.149.

Abstract

We have developed a method to calculate flow noninvasively in blood vessels using color Motion-mode (M-mode) and computer postprocessing. The velocity of each point in the cross-sectional area of the vessel was found from the color M-mode recording by correcting for angle both distances and velocities and by assuming a symmetrical circular velocity field. Volume flow was then found by integrating the velocity field at 5-ms intervals through the cardiac cycle. In a cardiovascular hydromechanical model, a correlation of 0.99 and p value of less than 0.001 were found between estimated and measured flow in the model (n = 8). In 20 healthy individuals, we made 31 investigations in the common carotid (CCA), internal carotid (ICA), and external carotid (ECA) artery, comparing flow in the CCA with the added flow in the ICA and ECA. The values (CCA versus ICA + ECA) correlated with r = 0.91 and p less than 0.01. Repeated investigations (n = 8) in one individual gave flow estimates of 495 +/- 50 ml/min in the CCA, 304 +/- 45 ml/min in the ICA, and 165 +/- 37 ml/min in the ECA (means +/- SD). This article shows that this system can make accurate estimation of blood flow to the brain noninvasively.

MeSH terms

  • Blood Flow Velocity
  • Carotid Arteries / physiology*
  • Color
  • Diagnosis, Computer-Assisted*
  • Microcomputers
  • Models, Cardiovascular
  • Models, Structural
  • Ultrasonics*