High incidence of defective ultrasound transducers in use in routine clinical practice

Eur J Echocardiogr. 2009 May;10(3):389-94. doi: 10.1093/ejechocard/jen295. Epub 2008 Oct 22.

Abstract

Aims: The objective was to evaluate the function of ultrasound transducers in use in routine clinical practice and thereby estimating the incidence of defective transducers.

Methods and results: The study comprised a one-time test of 676 transducers from 7 manufacturers which were in daily use in clinical departments at 32 hospitals. They were tested with the Sonora FirstCall Test System; 39.8% exhibited a transducer error. Delamination was detected in 26.5% and break in the cable was detected in 8.4% of the tested transducers. Errors originating from the piezoelectrical elements were unusual. Delamination and short circuit occurred without significant differences between transducers from all tested manufacturers, but the errors break in the cable, weak and dead element showed a statistically significant higher frequency in transducers from certain manufacturers.

Conclusion: The high error frequency and the risk for incorrect medical decisions when using a defective transducer indicate an urgent need for increased testing of the transducers in clinical departments.

Publication types

  • Case Reports
  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Child
  • Ductus Arteriosus, Patent / diagnostic imaging
  • Equipment Failure Analysis* / methods
  • Female
  • Humans
  • Quality Assurance, Health Care
  • Sensitivity and Specificity
  • Transducers / standards*
  • Ultrasonography, Doppler, Color / instrumentation*