Ion mobility-mass spectrometer interface for collisional activation of mobility separated ions

Anal Chem. 2009 Jan 15;81(2):618-24. doi: 10.1021/ac801919n.

Abstract

An ion mobility-mass spectrometer (IM-MS) interface is described that can be employed to perform collisional activation and/or collision-induced dissociation (CID) with good transmission of mobility separated ions to the MS analyzer. The IM-MS interface consists of a stacked-ring ion guide design, where the field strength and pressure ratio can be operated such that structural rearrangement reactions and/or CID are achieved as a function of the effective ion temperature. The ion dynamics and collisional activation processes in the IM-MS interface are described as a function of the ion-neutral collisions, ion kinetic energies, and effective ion temperature. The applicability of the IM-CID-MS methodology to studies of peptide ion fragmentation is illustrated using a series of model peptides.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Amino Acid Sequence
  • Ions / chemistry*
  • Peptides / chemistry
  • Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization / methods*

Substances

  • Ions
  • Peptides