Block circulant quasi-band matrix property for the SENSE unfolding in k-space and justification for GRAPPA

Annu Int Conf IEEE Eng Med Biol Soc. 2008:2008:1659-62. doi: 10.1109/IEMBS.2008.4649493.

Abstract

The k-space unfolding matrix for parallel imaging in MRI was examined and was found to have the structure of block circulant quasi-band, providing a rigorous mathematical justification for the interpolation kernel in the commonly used GRAPPA method. The optimal GRAPPA kernel size and dimension were closely related to sensitivity spectrum.

MeSH terms

  • Biomedical Engineering
  • Databases, Factual
  • Fourier Analysis
  • Heart / anatomy & histology
  • Humans
  • Image Processing, Computer-Assisted / statistics & numerical data*
  • Magnetic Resonance Imaging / statistics & numerical data*
  • Magnetic Resonance Imaging, Cine / statistics & numerical data
  • Models, Theoretical