Interlaboratory comparison of methodologies for the measurement of urinary 8-oxo-7,8-dihydro-2'-deoxyguanosine

Biomarkers. 2009 Mar;14(2):103-10. doi: 10.1080/13547500802706012.

Abstract

Urinary 8-oxo-7,8-dihydro-2'-deoxyguanosine (8-oxodGuo) is widely used as a marker of oxidative stress. Here we report the comparison of two, distinct chromatographic assays with an enzyme-linked immunosorbent assay (ELISA). The chromatographic assays displayed good agreement (r =:0.89, p < 0.0001), whereas there was markedly worse, albeit still significant, agreement with the ELISA (high-pressure liquid chromatography followed by gas chromatography (HPLC-GC/MS), r = 0.43; HPLC with electrochemical detection (HPLC-EC), r = 0.56; p < 0.0001). Mean values differed significantly between the chromatographic assays and the ELISA (HPLC-GC/MS 3.86, HPLC-EC 4.20, ELISA 18.70 ng mg(-1) creatinine; p < 0.0001). While it is reassuring to note good agreement between chromatographic assays, this study reveals significant short-comings in the ELISA, which brings into question its continued use in its present form.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • 8-Hydroxy-2'-Deoxyguanosine
  • Adult
  • Chromatography, High Pressure Liquid
  • Creatinine / urine
  • Deoxyguanosine / analogs & derivatives*
  • Deoxyguanosine / urine
  • Electrochemistry
  • Enzyme-Linked Immunosorbent Assay
  • Gas Chromatography-Mass Spectrometry
  • Humans
  • Laboratories*
  • Male
  • Middle Aged

Substances

  • 8-Hydroxy-2'-Deoxyguanosine
  • Creatinine
  • Deoxyguanosine