Surgery for complications of trans-catheter closure of atrial septal defects: a multi-institutional study from the European Congenital Heart Surgeons Association

Eur J Cardiothorac Surg. 2010 Jun;37(6):1285-90. doi: 10.1016/j.ejcts.2009.12.021. Epub 2010 Mar 28.

Abstract

Objective: This study aims to analyse the collective experience of participating European Congenital Heart Surgeons Association centres in the surgical management of complications resulting from trans-catheter closure of atrial septal defects (ASDs).

Methods: The records of all (n=56) patients, aged 3-70 years (median 18 years), who underwent surgery for complications of trans-catheter ASD closure in 19 participating institutions over a 10-year period (1997-2007) were retrospectively reviewed. Risk factors for surgical complications were sought. Surgical outcomes were compared with those reported for primary surgical ASD closure in the European Association of Cardio-thoracic Surgery Congenital Database.

Results: A wide range of ASD sizes (5-34mm) and devices of various types and sizes (range 12-60mm) were involved, including 13 devices less than 20mm. Complications leading to surgery included embolisation (n=29), thrombosis/thrombo-embolism/cerebral ischaemia or stroke (n=12), significant residual shunt (n=12), aortic or atrial perforation or erosion (n=9), haemopericardium with tamponade (n=5), aortic or mitral valve injury (n=2) and endocarditis (n=1). Surgery (39 early emergent and 17 late operations) involved device removal, repair of damaged structures and ASD closure. Late operations were needed 12 days to 8 years (median 3 years) after device implantation. There were three hospital deaths (mortality 5.4%). During the same time period, mortality for all 4453 surgical ASD closures reported in the European Association of Cardio-Thoracic Surgery Congenital Database was 0.36% (p=0.001).

Conclusions: Trans-catheter device closure of ASDs, even in cases when small devices are used, can lead to significant complications requiring surgical intervention. Once a complication leading to surgery occurs, mortality is significantly greater than that of primary surgical ASD closure. Major complications can occur late after device placement. Therefore, lifelong follow-up of patients in whom ASDs have been closed by devices is mandatory.

Publication types

  • Multicenter Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Cardiac Catheterization / adverse effects*
  • Cardiac Catheterization / methods
  • Child
  • Child, Preschool
  • Device Removal
  • Emergencies
  • Epidemiologic Methods
  • Europe / epidemiology
  • Female
  • Heart Septal Defects, Atrial / mortality
  • Heart Septal Defects, Atrial / surgery*
  • Humans
  • Long-Term Care
  • Male
  • Middle Aged
  • Minimally Invasive Surgical Procedures / adverse effects
  • Postoperative Complications / surgery
  • Septal Occluder Device / adverse effects*
  • Treatment Outcome