Contemporary use of embolic protection devices in saphenous vein graft interventions: Insights from the stenting of saphenous vein grafts trial

Catheter Cardiovasc Interv. 2010 Aug 1;76(2):263-9. doi: 10.1002/ccd.22438.

Abstract

Background: We sought to evaluate the contemporary use of embolic protection devices (EPDs) in saphenous vein graft (SVG) interventions.

Methods: We examined EPD use in the stenting of saphenous vein grafts (SOS) trial, in which 80 patients with 112 lesions in 88 SVGs were randomized to a bare metal stent (39 patients, 43 grafts, and 55 lesions) or paclitaxel-eluting stent (41 patients, 45 grafts, and 57 lesions).

Results: An EPD was used in 60 of 112 lesions (54%). A Filterwire (Boston Scientific) was used in 70% of EPD-treated lesions, Spider (ev3, Plymouth, Minnesota) in 12%, Proxis (St. Jude, Minneapolis, Minnesota) in 12%, and Guardwire (Medtronic, Santa Rosa, California) in 7%. Of the remaining 52 lesions, an EPD was not utilized in 13 lesions (25%) because the lesion was near the distal anastomosis, in 14 lesions (27%) because of an ostial location, in one lesion (2%) because of small SVG size, in two in-stent restenosis lesions (4%) because of low distal embolization risk, and in 22 lesions (42%) because of operator's preference even though use of an EPD was feasible. Procedural success was achieved in 77 patients (96%); in one patient a Filterwire was entrapped requiring emergency coronary bypass graft surgery and two patients had acute stent thrombosis.

Conclusion: In spite of their proven efficacy, EPDs were utilized in approximately half of SVG interventions in the SOS trial. Availability of a proximal protection device could allow protection of approximately 25% of unprotected lesions, yet operator discretion appears to be the major determinant of EPD use.

Trial registration: ClinicalTrials.gov NCT00247208.

Publication types

  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Aged
  • Angioplasty, Balloon, Coronary / adverse effects
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Coronary Angiography
  • Coronary Artery Bypass / adverse effects*
  • Drug-Eluting Stents
  • Embolism / etiology
  • Embolism / prevention & control*
  • Graft Occlusion, Vascular / diagnostic imaging
  • Graft Occlusion, Vascular / etiology
  • Graft Occlusion, Vascular / therapy*
  • Humans
  • Male
  • Metals
  • Middle Aged
  • Prosthesis Design
  • Saphenous Vein / diagnostic imaging
  • Saphenous Vein / transplantation*
  • Single-Blind Method
  • Stents*
  • Treatment Outcome
  • United States

Substances

  • Metals

Associated data

  • ClinicalTrials.gov/NCT00247208