Imaging near metal with a MAVRIC-SEMAC hybrid

Magn Reson Med. 2011 Jan;65(1):71-82. doi: 10.1002/mrm.22523.

Abstract

The recently developed multi-acquisition with variable resonance image combination (MAVRIC) and slice-encoding metal artifact correction (SEMAC) techniques can significantly reduce image artifacts commonly encountered near embedded metal hardware. These artifact reductions are enabled by applying alternative spectral and spatial-encoding schemes to conventional spin-echo imaging techniques. Here, the MAVRIC and SEMAC concepts are connected and discussed. The development of a hybrid technique that utilizes strengths of both methods is then introduced. The presented technique is shown capable of producing minimal artifact, high-resolution images near total joint replacements in a clinical setting.