Effect of limbal marking prior to laser ablation on the magnitude of cyclotorsional error

J Refract Surg. 2012 May;28(5):358-62. doi: 10.3928/1081597X-20120221-02. Epub 2012 Mar 12.

Abstract

Purpose: To evaluate the residual registration error after limbal-marking-based manual adjustment in cyclotorsional tracker-controlled laser refractive surgery.

Methods: Two hundred eyes undergoing custom surface ablation with the iVIS Suite (iVIS Technologies) were divided into limbal marked (marked) and non-limbal marked (unmarked) groups. Iris registration information was acquired preoperatively from all eyes. Preoperatively, the horizontal axis was recorded in the marked group for use in manual cyclotorsional alignment prior to surgical iris registration. During iris registration, the preoperative iris information was compared to the eye-tracker captured image. The magnitudes of the registration error angle and cyclotorsional movement during the subsequent laser ablation were recorded and analyzed.

Results: Mean magnitude of registration error angle (absolute value) was 1.82°±1.31° (range: 0.00° to 5.50°) and 2.90°±2.40° (range: 0.00° to 13.50°) for the marked and unmarked groups, respectively (P<.001). Mean magnitude of cyclotorsional movement during the laser ablation (absolute value) was 1.15°±1.34° (range: 0.00° to 7.00°) and 0.68°±0.97° (range: 0.00° to 6.00°) for the marked and unmarked groups, respectively (P=.005). Forty-six percent and 60% of eyes had registration error >2°, whereas 22% and 20% of eyes had cyclotorsional movement during ablation >2° in the marked and unmarked groups, respectively.

Conclusions: Limbal-marking-based manual alignment prior to laser ablation significantly reduced cyclotorsional registration error. However, residual registration misalignment and cyclotorsional movements remained during ablation.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Corneal Diseases / diagnosis*
  • Corneal Diseases / etiology
  • Humans
  • Iris / anatomy & histology*
  • Lasers, Excimer / therapeutic use*
  • Limbus Corneae / anatomy & histology*
  • Middle Aged
  • Refractive Errors / prevention & control
  • Torsion Abnormality / diagnosis*
  • Torsion Abnormality / etiology