Novel device (AirWave) to assess endotracheal tube migration: a pilot study

J Crit Care. 2013 Aug;28(4):535.e1-8. doi: 10.1016/j.jcrc.2012.10.015. Epub 2013 Feb 5.

Abstract

Introduction: Little is known about endotracheal tube (ETT) migration during routine care among critically ill patients. AirWave is a novel device that uses sonar waves to measure ETT migration and obstructions in real time. The aim of the present study is to assess the accuracy of the AirWave to evaluate ETT migration. In addition, we determined the degree of variation in ETT position and tested whether more pronounced migration occurs in specific clinical scenarios.

Methods: After institutional review board approval, we included mechanically ventilated patients from February 2012 to May 2012. A chest radiography (CXR) was obtained at baseline and 24 hours when clinically indicated. The ETT distance at the lips was recorded at baseline and every 4 hours. The AirWave system continuously recorded ETT position changes from baseline, and luminal obstructions.

Results: A total of 42 patients (age: 61 [SD ±13] years, men: 52%) were recruited. A total of 19 patients had measurements of ETT migration at 24 hours by the 3 methodologies used in this study. The mean (SD) of the ETT migration at 24 hours was +0.04 (1.2), -0.42 (0.7) and +0.34 (1.81) cm when measured by portable CXR, ETT distance at the teeth and AirWave device, respectively. Bland-Altman analysis of tube migration at 24 hours comparing the AirWave with CXR readings showed a bias of 0.1 cm with 95% limit of agreement of -3.8 and +4.3 cm. Comparison of tube migration at 24 hours determined by AirWave with ETT distance at the lips revealed a bias of -0.4 with 95% limit of agreement -3.7 to +3 cm, similar to the values observed between CXR and ETT distance at the lips (bias of -0.3 cm, 95% limit of agreement of -3.4 to +2.8 cm). Factors associated with ETT migration at 24 hours were ETT size and initial measurement from ETT tip to carina by portable CXR. AirWave detected in eight patients some degree of ETT obstruction (30% ± 9.6%) that resolved with prompt ETT catheter suction.

Conclusions: The AirWave may provide useful information regarding ETT migration and obstruction in real time.

Keywords: AirWave; Endotracheal tube; Migration; Obstruction.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Critical Illness*
  • Cross-Sectional Studies
  • Equipment Design
  • Female
  • Foreign-Body Migration / diagnosis*
  • Humans
  • Intubation, Intratracheal / instrumentation*
  • Male
  • Middle Aged
  • Monitoring, Physiologic / instrumentation*
  • Pilot Projects
  • Prospective Studies
  • Radiography, Thoracic
  • Respiration, Artificial
  • Sound*