A prototype retractor system designed to minimize ischemic brain retractor injury: initial observations

Surg Neurol. 1990 Sep;34(3):139-43. doi: 10.1016/0090-3019(90)90062-t.

Abstract

The authors have developed and patented a neurosurgical retractor system incorporating an infrared emitter and detector that allows detection of cerebral pulsations. Gentle contact with the surface of cat brains shows cerebral pulsations that correlate with arterial pulse as well as mechanical ventilation. The amplitude of cerebral pulsations decreases with higher retraction pressure and disappears at approximately 20 mmHg. The pressure on the surface of the brain decreased 50% in 5 minutes even though the position of the retractor was maintained constant. The authors postulate that monitoring cerebral pulsation may prove useful in clinical neurosurgery with respect to avoiding excessive retraction, which causes brain damage.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Brain Ischemia / etiology
  • Brain Ischemia / prevention & control*
  • Cats
  • Cerebrovascular Circulation / physiology
  • Intraoperative Complications / prevention & control*
  • Monitoring, Physiologic
  • Neurosurgery / instrumentation*
  • Surgical Instruments