Integrated laser-guided CT biopsy

Clin Imaging. 2013 Nov-Dec;37(6):1135-7. doi: 10.1016/j.clinimag.2013.08.006. Epub 2013 Sep 12.

Abstract

The diagnostic yield of computed tomography (CT)-guided biopsies is dependent on accurate needle insertion. A laser-assisted angle selection system was custom fabricated and used during a CT-guided lung biopsy. Off-target error was measured comparing standard methods to the laser method while pointing towards the target from the skin. The difference between the planned angle and selected angle using the laser-assisted system was 2°, improved from 12° with the standard method. Although yet to be confirmed, laser-assisted angle selection systems may improve the accuracy of needle placement, which may translate into improved outcomes for certain needle based procedures.

Keywords: Angle selection; Biopsy; CT guidance; Interventional; Laser guidance.

Publication types

  • Case Reports
  • Evaluation Study
  • Research Support, N.I.H., Intramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biopsy, Needle / instrumentation
  • Biopsy, Needle / methods*
  • Granulomatous Disease, Chronic / diagnostic imaging
  • Granulomatous Disease, Chronic / pathology
  • Humans
  • Image-Guided Biopsy / instrumentation
  • Image-Guided Biopsy / methods*
  • Lasers*
  • Lung / pathology*
  • Male
  • Middle Aged
  • Pneumonia / diagnostic imaging
  • Pneumonia / pathology
  • Recurrence
  • Tomography, X-Ray Computed / instrumentation
  • Tomography, X-Ray Computed / methods*