Low efficacy of cardioversion of persistent atrial fibrillation with the implantable cardioverter-defibrillator

Neth Heart J. 2013 Dec;21(12):548-53. doi: 10.1007/s12471-013-0474-z.

Abstract

Aims: Atrial fibrillation (AF) and heart failure are conditions that often coexist. Consequently, many patients with an implantable cardioverter-defibrillator (ICD) present with AF. We evaluated the effectiveness of internal cardioversion of AF in patients with an ICD.

Methods: Retrospectively, we included 27 consecutive ICD patients with persistent AF who underwent internal cardioversion using the ICD. When ICD cardioversion failed, external cardioversion was performed.

Results: Patients were predominantly male (89 %) with a mean (SD) age of 65 ± 9 years and left ventricular ejection fraction of 36 ± 17 %. Only nine (33 %) patients had successful internal cardioversion after one, two or three shocks. The remaining 18 patients underwent external cardioversion after they failed internal cardioversion, which resulted in sinus rhythm in all. A smaller left atrial volume (99 ± 36 ml vs. 146 ± 44 ml; p = 0.019), a longer right atrial cycle length (227 (186-255) vs. 169 (152-183) ms, p = 0.030), a shorter total AF history (2 (0-17) months vs. 40 (5-75) months, p = 0.025) and dual-coil ICD shock (75 % vs. 26 %, p = 0.093) were associated with successful ICD cardioversion.

Conclusion: Internal cardioversion of AF in ICD patients has a low success rate but may be attempted in those with small atria, a long right atrial fibrillatory cycle length and a short total AF history, especially when a dual-coil ICD is present. Otherwise, it seems reasonable to prefer external over internal cardioversion when it comes to termination of persistent AF.