Optimization of K-edge imaging for vulnerable plaques using gold nanoparticles and energy resolved photon counting detectors: a simulation study

Phys Med Biol. 2014 Jan 6;59(1):135-52. doi: 10.1088/0031-9155/59/1/135. Epub 2013 Dec 13.

Abstract

We investigated the effect of different imaging parameters, such as dose, beam energy, energy resolution and the number of energy bins, on the image quality of K-edge spectral computed tomography (CT) of gold nanoparticles (GNP) accumulated in an atherosclerotic plaque. A maximum likelihood technique was employed to estimate the concentration of GNP, which served as a targeted intravenous contrast material intended to detect the degree of the plaque's inflammation. The simulation studies used a single-slice parallel beam CT geometry with an x-ray beam energy ranging between 50 and 140 kVp. The synthetic phantoms included small (3 cm in diameter) cylinder and chest (33 × 24 cm(2)) phantoms, where both phantoms contained tissue, calcium and gold. In the simulation studies, GNP quantification and background (calcium and tissue) suppression tasks were pursued. The x-ray detection sensor was represented by an energy resolved photon counting detector (e.g., CdZnTe) with adjustable energy bins. Both ideal and more realistic (12% full width at half maximum (FWHM) energy resolution) implementations of the photon counting detector were simulated. The simulations were performed for the CdZnTe detector with a pixel pitch of 0.5-1 mm, which corresponds to a performance without significant charge sharing and cross-talk effects. The Rose model was employed to estimate the minimum detectable concentration of GNPs. A figure of merit (FOM) was used to optimize the x-ray beam energy (kVp) to achieve the highest signal-to-noise ratio with respect to the patient dose. As a result, the successful identification of gold and background suppression was demonstrated. The highest FOM was observed at the 125 kVp x-ray beam energy. The minimum detectable GNP concentration was determined to be approximately 1.06 µmol mL(-1) (0.21 mg mL(-1)) for an ideal detector and about 2.5 µmol mL(-1) (0.49 mg mL(-1)) for a more realistic (12% FWHM) detector. The studies show the optimal imaging parameters at the lowest patient dose using an energy resolved photon counting detector to image GNP in an atherosclerotic plaque.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Cadmium
  • Gold / chemistry*
  • Gold / metabolism*
  • Metal Nanoparticles*
  • Models, Theoretical
  • Phantoms, Imaging
  • Photons*
  • Plaque, Atherosclerotic / diagnostic imaging*
  • Polymethyl Methacrylate
  • Scintillation Counting / methods*
  • Signal-To-Noise Ratio
  • Tellurium
  • Tomography, X-Ray Computed / methods*
  • Zinc

Substances

  • CdZnTe
  • Cadmium
  • Gold
  • Polymethyl Methacrylate
  • Zinc
  • Tellurium