Packaging and modular assembly of large-area and fine-pitch 2-D ultrasonic transducer arrays

IEEE Trans Ultrason Ferroelectr Freq Control. 2013 Jul;60(7):1356-75. doi: 10.1109/TUFFC.2013.2709.

Abstract

A promising transducer architecture for largearea arrays employs 2-D capacitive micromachined ultrasound transducer (CMUT) devices with backside trench-frame pillar interconnects. Reconfigurable array (RA) application-specified integrated circuits (ASICs) can provide efficient interfacing between these high-element-count transducer arrays and standard ultrasound systems. Standard electronic assembly techniques such as flip-chip and ball grid array (BGA) attachment, along with organic laminate substrate carriers, can be leveraged to create large-area arrays composed of tiled modules of CMUT chips and interface ASICs. A large-scale, fully populated and integrated 2-D CMUT array with 32 by 192 elements was developed and demonstrates the feasibility of these techniques to yield future large-area arrays. This study demonstrates a flexible and reliable integration approach by successfully combining a simple under-bump metallization (UBM) process and a stacked CMUT/interposer/ASIC module architecture. The results show high shear strength of the UBM (26.5 g for 70-μm balls), high interconnect yield, and excellent CMUT resonance uniformity (s = 0.02 MHz). A multi-row linear array was constructed using the new CMUT/interposer/ASIC process using acoustically active trench-frame CMUT devices and mechanical/ nonfunctional Si backside ASICs. Imaging results with the completed probe assembly demonstrate a functioning device based on the modular assembly architecture.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Equipment Design
  • Phantoms, Imaging
  • Transducers*
  • Ultrasonography / instrumentation*