Owing to its intrinsic characteristics, spatiotemporally encoded (SPEN) imaging is less sensitive to adverse effects due to field inhomogeneity in comparison with echo planar imaging, a feature highly desired for functional, diffusion, and real-time MRI. However, the quality of images obtained with SPEN MRI is still degraded by geometric distortions when field inhomogeneity exists. In this study, a single-shot biaxial SPEN (bi-SPEN) pulse sequence is implemented, utilizing a 90° and a 180° chirp pulse incorporated with two orthogonal gradients. A referenceless geometric-distortion correction based on the single-shot bi-SPEN sequence is then proposed. The distorted image acquired with the single-shot bi-SPEN sequence is corrected by iterative super-resolved reconstruction involving the field gradients estimated from a field map, which in turn is obtained from its own super-resolved data after a phase-unwrapping procedure without additional scans. In addition, the distortion correction method is applied to improve the quality of the multiple region-of-interest images obtained with single-shot bi-SPEN sequence.
Keywords: Biaxial spatiotemporal encoding; Iterative reconstruction; Referenceless distortion correction; Region-of-interest imaging; Super-resolved reconstruction; Ultrafast MRI.