Triple taper stem design shows promising fixation and bone remodelling characteristics: radiostereometric analysis in a randomised controlled trial

Bone Joint J. 2015 Jun;97-B(6):755-61. doi: 10.1302/0301-620X.97B6.34736.

Abstract

Cemented femoral stems with force closed fixation designs have shown good clinical results despite high early subsidence. A new triple-tapered stem in this category (C-stem AMT) was introduced in 2005. This study compares this new stem with an established stem of similar design (Exeter) in terms of migration (as measured using radiostereometric analysis), peri-prosthetic bone remodelling (measured using dual energy x-ray densitometry, DXA), Oxford Hip Score, and plain radiographs. A total of 70 patients (70 hips) with a mean age of 66 years (53 to 78) were followed for two years. Owing to missing data of miscellaneous reasons, the final analysis represents data from 51 (RSA) and 65 (DXA) patients. Both stems showed a typical pattern of migration: Subsidence and retroversion that primarily occurred during the first three months. C-stem AMT subsided less during the first three months (p = 0.01), before stabilising at a subsidence rate similar to the Exeter stem from years one to two. The rate of migration into retroversion was slightly higher for C-stem AMT during the second year (p = 0.03). Whilst there were slight differences in movement patterns between the stems, the C-stem AMT exhibits good early clinical outcomes and displays a pattern of migration and bone remodelling that predicts good clinical performance.

Keywords: Cemented stem; Hip; RSA; polished tapered stem.

Publication types

  • Comparative Study
  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Bone Remodeling*
  • Female
  • Femur
  • Hip Prosthesis*
  • Humans
  • Male
  • Middle Aged
  • Prosthesis Design*
  • Radiostereometric Analysis
  • Stress, Mechanical