Sensor-Based Electromagnetic Navigation (Mediguide®): How Accurate Is It? A Phantom Model Study

J Cardiovasc Electrophysiol. 2015 Oct;26(10):1140-5. doi: 10.1111/jce.12741. Epub 2015 Jul 27.

Abstract

Background: Data about localization reproducibility as well as spatial and visual accuracy of the new MediGuide® sensor-based electroanatomic navigation technology are scarce. We therefore sought to quantify these parameters based on phantom experiments.

Methods and results: A realistic heart phantom was generated in a 3D-Printer. A CT scan was performed on the phantom. The phantom itself served as ground-truth reference to ensure exact and reproducible catheter placement. A MediGuide® catheter was repeatedly tagged at selected positions to assess accuracy of point localization. The catheter was also used to acquire a MediGuide®-scaled geometry in the EnSite Velocity® electroanatomic mapping system. The acquired geometries (MediGuide®-scaled and EnSite Velocity®-scaled) were compared to a CT segmentation of the phantom to quantify concordance. Distances between landmarks were measured in the EnSite Velocity®- and MediGuide®-scaled geometry and the CT dataset for Bland-Altman comparison. The visualization of virtual MediGuide® catheter tips was compared to their corresponding representation on fluoroscopic cine-loops. Point localization accuracy was 0.5 ± 0.3 mm for MediGuide® and 1.4 ± 0.7 mm for EnSite Velocity®. The 3D accuracy of the geometries was 1.1 ± 1.4 mm (MediGuide®-scaled) and 3.2 ± 1.6 mm (not MediGuide®-scaled). The offset between virtual MediGuide® catheter visualization and catheter representation on corresponding fluoroscopic cine-loops was 0.4 ± 0.1 mm.

Conclusions: The MediGuide® system shows a very high level of accuracy regarding localization reproducibility as well as spatial and visual accuracy, which can be ascribed to the magnetic field localization technology. The observed offsets between the geometry visualization and the real phantom are below a clinically relevant threshold.

Keywords: 3D mapping system; catheter ablation; catheter navigation; image integration.

Publication types

  • Evaluation Study

MeSH terms

  • Body Surface Potential Mapping / instrumentation*
  • Body Surface Potential Mapping / methods
  • Cardiovascular Surgical Procedures / instrumentation*
  • Cardiovascular Surgical Procedures / methods
  • Catheter Ablation / instrumentation*
  • Catheter Ablation / methods
  • Electromagnetic Fields
  • Equipment Design
  • Equipment Failure Analysis
  • Humans
  • Magnetics / instrumentation*
  • Phantoms, Imaging
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Surgery, Computer-Assisted / instrumentation*
  • Surgery, Computer-Assisted / methods