Simplifying sample preparation using fabric phase sorptive extraction technique for the determination of benzodiazepines in blood serum by high-performance liquid chromatography

Biomed Chromatogr. 2016 Jun;30(6):829-36. doi: 10.1002/bmc.3615. Epub 2015 Oct 15.

Abstract

Fabric phase sorptive extraction (FPSE), a recently introduced novel sample preparation technology, has been evaluated for the extraction of benzodiazepines from human blood serum. FPSE utilizes a flexible fabric surface as the substrate platform for creating sol-gel hybrid organic-inorganic sorbent coatings. FPSE media can be introduced directly into the sample containing the target analyte(s), requiring no need for prior sample pretreatment or clean-up. Benzodiazepines were selected as model analytes because they represent one of the most widely used therapeutic drugs in psychiatry and are also amongst the most frequently encountered drugs in forensic toxicology. The chromatographic separation of target analytes was performed on a LiChroCART-LiChrospher®100 RP-18e (5 µm, 250 × 4 mm) analytical column, operated at room temperature. Ternary gradient elution was applied with a mobile phase that consisted of acetonitrile, methanol and ammonium acetate (0.05 M), which was delivered at a flow rate of 1.0 mL/min. Diode array detection was performed with monitoring at 240 nm. FPSE was performed using cellulose fabric extraction media coated with sol-gel poly(ethylene glycol) (sol-gel PEG). Absolute recovery values in the equilibrium state for the examined benzodiazepines were found to be 27% for bromazepam, 63% for lorazepam, 42 % for diazepam and 39% for alprazolam. Copyright © 2015 John Wiley & Sons, Ltd.

Keywords: Fabric Phase Sorptive Extraction; Sample Preparation; Sorptive Extraction; benzodiazepines; blood serum.

Publication types

  • Validation Study

MeSH terms

  • Benzodiazepines / blood*
  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Reference Standards
  • Reproducibility of Results

Substances

  • Benzodiazepines