Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories

Phys Med. 2016 Jan;32(1):42-51. doi: 10.1016/j.ejmp.2015.09.008. Epub 2015 Oct 23.

Abstract

A comparison of calibration results and procedures in terms of air kerma length product, PKL, and air kerma, K, was conducted between eight dosimetry laboratories. A pencil-type ionization chamber (IC), generally used for computed tomography dose measurements, was calibrated according to three calibration methods, while its residual signal and other characteristics (sensitivity profile, active length) were assessed. The results showed that the "partial irradiation method" is the preferred method for the pencil-type IC calibration in terms of PKL and it could be applied by the calibration laboratories successfully. Most of the participating laboratories achieved high level of agreement (>99%) for both dosimetry quantities (PKL and K). Estimated relative standard uncertainties of comparison results vary among laboratories from 0.34% to 2.32% depending on the quantity, beam quality and calibration method applied. Detailed analysis of the assigned uncertainties have been presented and discussed.

Keywords: Air kerma length product; CT dosimetry; CT pencil type ionization chambers; Calibration; Comparison.

Publication types

  • Comparative Study

MeSH terms

  • Air
  • Calibration
  • Europe
  • International Cooperation
  • Ions*
  • Laboratories / standards*
  • Models, Statistical
  • Observer Variation
  • Radiation Dosage
  • Radiometry / methods*
  • Reproducibility of Results
  • Tomography, X-Ray Computed / methods*