Efficacy of a balloon-expandable vascular access system in transfemoral TAVI patients

Catheter Cardiovasc Interv. 2016 Dec;88(7):1145-1152. doi: 10.1002/ccd.26514. Epub 2016 Mar 31.

Abstract

Background: Vascular complications (VC) are a serious and frequent complication of transfemoral transcatheter aortic valve implantation (TAVI) and result in increased morbidity and mortality. It has been suggested that newly developed vascular sheaths may increase the ability to perform transfemoral TAVI in patients with normal and access-limiting peripheral artery disease (PAD) and reduce vascular complications.

Aims: We sought to assess the safety and efficacy of the 19 French (F) SoloPath balloon-expandable transfemoral vascular access system in patients who underwent transfemoral TAVI at our center between 2011 and 2014.

Methods and results: Single-center retrospective study of 90 patients who underwent transfemoral TAVI with the use of the SoloPath sheath. Patients were categorized into two groups according to a sheath to femoral artery ratio (SFAR) of less than or equal to 1.05, or greater than 1.05. Overall, the incidence of major bleeding complications was low, 4.4%. No significant differences were found in technical or procedural success rates (100% in both groups and 100% vs. 91.3; P = 0.09; respectively), total vascular complications (20.8 vs. 21.7; P = 0.92) or total bleeding complications (20.8 vs. 30.4; P = 0.36 between those with SFAR greater or less than 1.05.

Conclusions: The use of the SoloPath balloon-expandable sheath is feasible and safe even in patients with SFAR > 1.05, showing no increased vascular or bleeding complications compared to patients with larger vascular access. © 2016 Wiley Periodicals, Inc.

Keywords: Structural Heart Disease Intervention (SHDI); Transcatheter Valve Implantation (TVI); Vascular Complications (VCOM).

MeSH terms

  • Aged
  • Aged, 80 and over
  • Angioplasty, Balloon / adverse effects
  • Angioplasty, Balloon / instrumentation*
  • Aortic Valve* / diagnostic imaging
  • Aortic Valve* / physiopathology
  • Cardiac Catheterization* / adverse effects
  • Cardiac Catheterization* / methods
  • Catheterization, Peripheral / adverse effects
  • Catheterization, Peripheral / instrumentation*
  • Catheterization, Peripheral / methods
  • Equipment Design
  • Feasibility Studies
  • Female
  • Femoral Artery* / diagnostic imaging
  • Femoral Artery* / injuries
  • Femoral Artery* / physiopathology
  • Heart Valve Diseases / complications
  • Heart Valve Diseases / diagnostic imaging
  • Heart Valve Diseases / physiopathology
  • Heart Valve Diseases / therapy*
  • Heart Valve Prosthesis Implantation* / adverse effects
  • Heart Valve Prosthesis Implantation* / methods
  • Hemorrhage / etiology
  • Humans
  • Male
  • Peripheral Arterial Disease / complications
  • Peripheral Arterial Disease / diagnostic imaging
  • Peripheral Arterial Disease / physiopathology
  • Peripheral Arterial Disease / therapy*
  • Punctures
  • Quebec
  • Retrospective Studies
  • Risk Factors
  • Treatment Outcome
  • Vascular Access Devices*
  • Vascular System Injuries / etiology