The Effect of Microcirculatory Flow on Oscillating Gradient Diffusion MRI and Diffusion Encoding with Dual-Frequency Orthogonal Gradients (DEFOG)

Magn Reson Med. 2017 Apr;77(4):1583-1592. doi: 10.1002/mrm.26242. Epub 2016 Apr 15.

Abstract

Purpose: We investigated the effect of microcirculatory flow on oscillating gradient spin echo (OGSE) diffusion MRI at low b-values and developed a diffusion preparation method called diffusion encoding with dual-frequency orthogonal gradients (DEFOG) to suppress the effect.

Methods: Compared to conventional OGSE sequences, DEFOG adds a pulsed gradient that is orthogonal to the oscillating gradient and has a moderate diffusion weighting (e.g., 300 s/mm2 ). In vivo MRI data were acquired from adult mouse brains (n = 5) on an 11.7 Tesla scanner, with diffusion times from 23.2 to 0.83 ms and b-values from 50 to 700 s/mm2 .

Results: Apparent diffusion coefficients (ADCs) measured using a conventional OGSE sequence at low b-values (< 200 mm2 /s) were significantly higher than those measured at moderate b-values (> 300 mm2 /s), potentially due to contributions from microcirculatory flow. In comparison, OGSE ADCs measured using the DEFOG method at low b-values were comparable to those measured at moderate b-values. The effect of microcirculatory flow on diffusion signals was diffusion time-dependent, and this dependency may reflect the capillary geometry and blood flow velocity in the mouse cortex.

Conclusion: Microcirculatory flow affects OGSE diffusion MRI measurements at low b-values, and this effect can be suppressed using the DEFOG method. Magn Reson Med 77:1583-1592, 2017. © 2016 International Society for Magnetic Resonance in Medicine.

Keywords: diffusion MRI; dual-frequency orthogonal gradients; microcirculatory flow; mouse brain; oscillating gradient; pseudo-diffusion.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Animals
  • Artifacts*
  • Blood Flow Velocity / physiology*
  • Diffusion Magnetic Resonance Imaging / methods*
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods
  • Male
  • Mice
  • Mice, Inbred C57BL
  • Microcirculation / physiology*
  • Oscillometry / methods
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Signal Processing, Computer-Assisted*