Analysis of acoustic impedance matching in dual-band ultrasound transducers

J Acoust Soc Am. 2017 Feb;141(2):1170. doi: 10.1121/1.4976096.

Abstract

Dual-frequency band probes are needed for ultrasound (US) reverberation suppression and are useful for image-guided US therapy. A challenge is to design transducer stacks that achieve high bandwidth and efficiency at both operating frequencies when the frequencies are widely separated with a frequency ratio ∼6:1-20:1. This paper studies the loading and backing conditions of transducers in such stacks. Three stack configurations are presented and analyzed using one-dimensional models. It is shown that a configuration with three layers of material separating the transducers is favorable, as it reduces high frequency ringing by ∼20 dB compared to other designs, and matches the low frequency (LF) transducer to the load at a lower frequency. In some cases, the LF load matching is governed by a simple mass-spring interaction in spite of having a complicated matching structure. The proposed design should yield improved performance of reverberation suppression algorithms. Its suitability for reduction of probe heating, also in single-band probes, should be investigated.