Comparison of Cone-Beam Tomography and Cross-Sectional Imaging for Volumetric and Dosimetric Calculations in Resin Yttrium-90 Radioembolization

Cardiovasc Intervent Radiol. 2018 Dec;41(12):1857-1866. doi: 10.1007/s00270-018-2030-0. Epub 2018 Jul 13.

Abstract

Purpose: To compare the use of cone-beam computed tomography versus contrast-enhanced computed tomography (CT) and magnetic resonance imaging (MRI) in the calculation of liver volume and planned dose for yttrium-90 radioembolization.

Materials and methods: The study retrospectively assessed 47 consecutive patients who underwent resin Y-90 radioembolization consecutively over a 2-year period at a single center. Volume calculation software was used to determine perfused lobar liver volumes from cone-beam CT (CBCT) images obtained during mapping angiography. CBCT-derived volumes were compared with perfused lobar volume derived from contrast-enhanced CT and MRI. Nominal activities as determined by the SIR-Spheres Microspheres Activity Calculator were similarly calculated and compared using both CBCT and conventionally acquired volumes.

Results: A total of 82 hepatic lobes were assessed in 47 patients. The mean percentage difference between combined CT-MRI- and CBCT-derived calculated lobar volumes was 25.3% (p = 0.994). The mean percentage difference in calculated dose between the two methods was 21.8 ± 24.6% (p = 0.42). Combined left and right lobar CT-derived dose difference was less than 10% in 22 lobes, between 10 and 25% in 20 lobes, between 25 and 50% in 13 lobes and greater than 50% in 5 lobes. Combined left and right lobar MRI-derived dose difference was less than 10% in 11 lobes, between 10 and 25% in 7 lobes, between 25 and 50% in 2 lobes and greater than 50% in 1 lobe.

Conclusions: Although volume measurements derived from CT/MRI did not differ significantly from those derived from CBCT, variability between the two methods led to large and unexpected differences in calculated dose.

Keywords: CBCT; Cone-beam CT; Dosimetry; Radioembolization; Y90.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Brachytherapy / methods*
  • Cone-Beam Computed Tomography / methods*
  • Contrast Media
  • Female
  • Humans
  • Image Enhancement / methods
  • Liver / diagnostic imaging
  • Liver / pathology
  • Liver / radiation effects
  • Liver Neoplasms / diagnostic imaging
  • Liver Neoplasms / pathology
  • Liver Neoplasms / radiotherapy*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Multimodal Imaging / methods
  • Organ Size
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Retrospective Studies
  • Tomography, X-Ray Computed / methods
  • Yttrium Radioisotopes / therapeutic use*

Substances

  • Contrast Media
  • Yttrium Radioisotopes
  • Yttrium-90