Magn Reson Med. 2019 Jan;81(1):57-68. doi: 10.1002/mrm.27440. Epub 2018 Aug 14.

Abstract

Purpose: To introduce a novel MR pulse sequence, termed Looping Star, for fast, robust, and yet quiet, 3D radial multi-gradient echo T2* MR imaging.

Methods: The Looping Star pulse sequence is based on the 3D radial Rotating Ultra-Fast Imaging Sequence (RUFIS) extended by a time-multiplexed gradient-refocusing mechanism. First, multiple magnetic coherences are excited, which are subsequently gradient-refocused in form of a looping k-space trajectory. Accordingly, Looping Star captures an initial FID image followed by gradient echo images at equidistant echo times.

Results: Looping Star was demonstrated in phantom and in vivo volunteer experiments for 3D, high resolution T2* weighted imaging, T2* mapping, and quantitative susceptibility mapping (QSM). The method is fast, quiet, and robust against imperfections including Eddy currents, motion, and geometric distortions. When applied to a motor task fMRI experiment a BOLD sensitivity of 5% was achieved at minimal acoustic noise (i.e. 2.7 dB(A) above ambient noise) and with images congruent to other anatomical scans.

Conclusions: Looping Star imaging provides new and exciting opportunities for fast, robust and yet quiet T2* MR imaging. Potential applications include T2*-weighted imaging, T2* mapping, QSM, and fMRI.

Keywords: BURST; Looping Star; QSM; Quiet; RUFIS; Silent; T2*; UTE; fMRI.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acoustics
  • Algorithms
  • Brain / diagnostic imaging*
  • Calibration
  • Humans
  • Image Enhancement / methods
  • Image Interpretation, Computer-Assisted / methods
  • Image Processing, Computer-Assisted
  • Imaging, Three-Dimensional / methods
  • Magnetic Resonance Imaging / methods*
  • Pattern Recognition, Automated
  • Phantoms, Imaging
  • Reproducibility of Results