Biocompatibility and Immunotoxicology of the Preclinical Implantation of a Collagen-based Artificial Dermal Regeneration Matrix

Biomed Environ Sci. 2018 Nov;31(11):829-842. doi: 10.3967/bes2018.110.

Abstract

Objective: Graft rejection, with the possibility of a violent immune response, may be severe and life threatening. Our aim was to thoroughly investigate the biocompatibility and immunotoxicology of collagen-based dermal matrix (DM) before assessment in clinical trials.

Methods: DM was subcutaneously implanted in BALB/c mice in two doses to induce a potential immune response. The spleen and lymph nodes were assessed for shape, cell number, cell phenotype via flow cytometry, cell activation via CCK8 kit, Annexin V kit, and Ki67 immunostaining. Serum samples were used to measure antibody concentration by enzyme-linked immunosorbent assay. Local inflammation was analyzed by histology and immunohistochemistry staining. Data analysis was performed by one-way ANOVA and non-parametric tests.

Results: Our data illustrate that the spleen and lymph node sizes were similar between the negative control mice and mice implanted with DM. However, in the high-dose DM (DM-H) group, the total cell populations in the spleen and lymph nodes, T cells and B cells in the spleen had slight increases in prophase, and the low-dose DM (DM-L) group did not display gross abnormities. Moreover, DM-H initiated moderate cell activation and proliferation in the early phase post-immunization, whereas DM-L did not. Neither DM-H nor DM-L implantation noticeably increased IgM and IgG serum concentrations. Examination of the local cellular response revealed only benign cell infiltration and TNF-α expression in slides of DM in the early phase.

Conclusion: Overall, DM-H may have induced a benign temporary acute immune response post-implantation, whereas DM-L had quite low immunogenicity. Thus, this DM can be regarded as a safe product.

Keywords: Collagen; Flow cytometry; Immunogenicity; Lymphocytes.

MeSH terms

  • Animals
  • Biocompatible Materials / adverse effects*
  • Biocompatible Materials / analysis
  • Collagen / adverse effects
  • Collagen / immunology*
  • Dermis / immunology
  • Dermis / surgery*
  • Female
  • Flow Cytometry
  • Immunity, Cellular
  • Lymph Nodes / immunology
  • Mice
  • Mice, Inbred BALB C
  • Prostheses and Implants / adverse effects*
  • Spleen / immunology

Substances

  • Biocompatible Materials
  • Collagen