Background: The response rate to cardiac resynchronization therapy (CRT) may be improved if echocardiographic-derived parameters are used to guide the left ventricular (LV) lead deployment. Tools to visually integrate deformation imaging and fluoroscopy to take advantage of the combined information are lacking.
Methods: An image fusion tool for echo-guided LV lead placement in CRT was developed. A personalized average 3D cardiac model aided visualization of patient-specific LV function in fluoroscopy. A set of coronary venography-derived landmarks facilitated registration of the 3D model with fluoroscopy into a single multimodality image. The fusion was both performed and analyzed retrospectively in 30 cases. Baseline time-to-peak values from echocardiography speckle-tracking radial strain traces were color-coded onto the fused LV. LV segments with suspected scar tissue were excluded by cardiac magnetic resonance imaging. The postoperative augmented image was used to investigate: (a) registration accuracy and (b) agreement between LV pacing lead location, echo-defined target segments, and CRT response.
Results: Registration time (264 ± 25 seconds) and accuracy (4.3 ± 2.3 mm) were found clinically acceptable. A good agreement between pacing location and echo-suggested segments was found in 20 (out of 21) CRT responders. Perioperative integration of the proposed workflow was successfully tested in 2 patients. No additional radiation, compared with the existing workflow, was required.
Conclusions: The fusion tool facilitates understanding of the spatial relationship between the coronary veins and the LV function and may help targeted LV lead delivery.
Keywords: cardiac resynchronization therapy; echocardiography; fluoroscopy; image fusion; image-guided intervention.
© 2019 Wiley Periodicals, Inc.