Reproducibility and Variability of Protein Analytes Measured Using a Multiplexed Modified Aptamer Assay

J Appl Lab Med. 2019 Jul;4(1):30-39. doi: 10.1373/jalm.2018.027086. Epub 2019 Jan 22.

Abstract

Background: There is growing interest in the use of multiplexed aptamer-based assays for large-scale proteomic studies. However, the analytic, short- and long-term variation of the measured proteins is largely uncharacterized.

Methods: We quantified 4001 plasma protein analytes from 42 participants in the Atherosclerosis Risk in Communities (ARIC) Study in split samples and at multiple visits using a multiplexed modified aptamer assay. We calculated the CV, Spearman correlation, and intraclass correlation (ICC) between split samples and evaluated the short-term (4-9 weeks) and long-term (approximately 20 years) variability using paired t-tests with log-transformed protein concentrations and Bonferroni-corrected significance thresholds. We performed principal component (PC) analysis of protein analyte concentrations and evaluated their associations with age, sex, race, and estimated glomerular filtration rate (eGFR).

Results: The mean baseline age was 57 years at the first visit, 43% of participants were male and 57% were white. Among 3693 protein analytes that passed quality control, half (n = 1846) had CVs < 5.0%, Spearman correlations > 0.89, and ICCs > 0.96 among the split samples. Over the short term, only 1 analyte had a statistically significant difference between the 2 time points, whereas, over approximately 20 years, 866 analytes (23.4%) had statistically significant differences (P < 1.4 × 10-5, 681 increased, 185 decreased). PC1 had high correlations with age (-0.73) and eGFR (0.60). PC2 had moderate correlation with male sex (0.18) and white race (0.31).

Conclusions: Multiplexed modified aptamer technology can assay thousands of proteins with excellent precision. Our results support the potential for large-scale studies of the plasma proteome over the lifespan.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Atherosclerosis / blood*
  • Atherosclerosis / diagnosis
  • Blood Proteins / analysis*
  • Equipment Design
  • Female
  • Glomerular Filtration Rate / physiology
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Proteomics / instrumentation*
  • Reproducibility of Results

Substances

  • Blood Proteins