A versatile Johansson-type tender x-ray emission spectrometer

Rev Sci Instrum. 2020 Mar 1;91(3):033101. doi: 10.1063/1.5121853.

Abstract

We present a high energy resolution x-ray spectrometer for the tender x-ray regime (1.6-5.0 keV) that was designed and operated at Stanford Synchrotron Radiation Lightsource. The instrument is developed on a Rowland geometry (500 mm of radius) using cylindrically bent Johansson analyzers and a position sensitive detector. By placing the sample inside the Rowland circle, the spectrometer operates in an energy-dispersive mode with a subnatural line-width energy resolution (∼0.32 eV at 2400 eV), even when an extended incident x-ray beam is used across a wide range of diffraction angles (∼30° to 65°). The spectrometer is enclosed in a vacuum chamber, and a sample chamber with independent ambient conditions is introduced to enable a versatile and fast-access sample environment (e.g., solid/gas/liquid samples, in situ cells, and radioactive materials). The design, capabilities, and performance are presented and discussed.