Abbreviated on-treatment CBCT using roughness penalized mono-energization of kV-MV data and a multi-layer MV imager

Phys Med Biol. 2021 Jun 21;66(13). doi: 10.1088/1361-6560/abddd2.

Abstract

Simultaneous acquisition of cone beam CT (CBCT) projections using both the kV and MV imagers of an image guided radiotherapy system reduces set-up scan times-a benefit to lung cancer radiation oncology patients-but increases noise in the 3D reconstruction. In this article, we present a kV-MV scan time reduction technique that uses two noise-reducing measures to achieve superior performance. The first is a high-DQE multi-layer MV imager prototype. The second is a beam hardening correction algorithm which combines poly-energetic modeling with edge-preserving, regularized smoothing of the projections. Performance was tested in real acquisitions of the Catphan 604 and a thorax phantom. Percent noise was quantified from voxel values in a soft tissue volume of interest (VOI) while edge blur was quantified from a VOI straddling a boundary between air and soft material. Comparisons in noise/resolution performance trade-off were made between our proposed approach, a dose-equivalent kV-only scan, and a kV-MV reconstruction technique previously published by Yinet al(2005Med. Phys.329). The proposed technique demonstrated lower noise as a function of spatial resolution than the baseline kV-MV method, notably a 50% noise reduction at typical edge blur levels. Our proposed method also exhibited fainter non-uniformity artifacts and in some cases superior contrast. Overall, we find that the combination of a multi-layer MV imager, acquiring at a LINAC source energy of 2.5 MV, and a denoised beam hardening correction algorithm enables noise, resolution, and dose performance comparable to standard kV-imager only set-up CBCT, but with nearly half the gantry rotation time.

Keywords: cone beam CT; kV‐MV; poly-energetic; radiation therapy; radiotherapy; scan time reduction.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Cone-Beam Computed Tomography
  • Humans
  • Particle Accelerators
  • Phantoms, Imaging
  • Radiotherapy, Image-Guided*
  • Spiral Cone-Beam Computed Tomography*