Sex-Based Differences in Outcomes With Percutaneous Transcatheter Repair of Mitral Regurgitation With the MitraClip System: Transcatheter Valve Therapy Registry From 2011 to 2017

Circ Cardiovasc Interv. 2021 Nov;14(11):e009374. doi: 10.1161/CIRCINTERVENTIONS.120.009374. Epub 2021 Nov 16.

Abstract

Background: Women have a higher rate of adverse events after mitral valve surgery. We sought to evaluate whether outcomes after transcatheter edge-to-edge repair intervention by sex have similar trends to mitral valve surgery.

Methods: The primary outcome was 1-year major adverse events defined as a composite of all-cause mortality, stroke, and any bleeding in the overall study cohort. Patients who underwent transcatheter edge-to-edge repair for mitral regurgitation with the MitraClip system in the Society of Thoracic Surgery/American College of Cardiology Transcatheter Valve Therapy registry were evaluated. Linked administrative claims from the Centers for Medicare and Medicaid Services were used to evaluate 1-year clinical outcomes. Associations between sex and outcomes were evaluated using a multivariable logistic regression model for in-hospital outcomes and Cox model for 1-year outcomes.

Results: From November 2013 to March 2017, 5295 patients, 47.6% (n=2523) of whom were female, underwent transcatheter edge-to-edge repair. Females were less likely to have >1 clip implanted (P<0.001) and had a lower adjusted odds ratio of device success (adjusted odds ratio, 0.78 [95% CI, 0.67-0.90]), driven by lower odds of residual mitral gradient <5 mm Hg (adjusted odds ratio, 0.54 [CI, 0.46-0.63]) when compared with males. At 1-year follow-up, the primary outcome did not differ by sex. Female sex was associated with lower adjusted 1-year risk of all-cause mortality (adjusted hazard ratio, 0.80 [CI, 0.68-0.94]), but the adjusted 1-year risk of stroke and any bleeding did not differ by sex.

Conclusions: No difference in composite outcome of all-cause mortality, stroke, and any bleeding was observed between females and males. Adjusted 1-year all-cause mortality was lower in females compared with males.

Keywords: follow-up studies; mitral valve; mortality; sex; transcatheter mitral valve repair.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Aged
  • Female
  • Heart Valve Prosthesis Implantation* / adverse effects
  • Humans
  • Male
  • Medicare
  • Mitral Valve Insufficiency* / diagnostic imaging
  • Mitral Valve Insufficiency* / epidemiology
  • Mitral Valve Insufficiency* / surgery
  • Registries
  • Treatment Outcome
  • United States / epidemiology