In Vitro Mitral Valve Model with Unrestricted Ventricular Access: Using Vacuum to Close the Valve and Enable Static Trans-Mitral Pressure

J Cardiovasc Transl Res. 2022 Aug;15(4):845-854. doi: 10.1007/s12265-021-10199-5. Epub 2022 Jan 6.

Abstract

Current in vitro models of the left heart establish the pressure difference required to close the mitral valve by sealing and pressurizing the ventricular side of the valve, limiting important access to the subvalvular apparatus. This paper describes and evaluates a system that establishes physiological pressure differences across the valve using vacuum on the atrial side. The subvalvular apparatus is open to atmospheric pressure and accessible by tools and sensors, establishing a novel technique for experimentation on atrioventricular valves. Porcine mitral valves were excised and closed by vacuum within the atrial chamber. Images were used to document and analyze closure of the leaflets. Papillary muscle force and regurgitant flow rate were measured to be 4.07 N at 120 mmHg and approximately 12.1 ml/s respectively, both of which are within clinically relevant ranges. The relative ease of these measurements demonstrates the usefulness of improved ventricular access at peak pressure/force closure.

Keywords: Chordae tendineae; In vitro; Mitral annulus; Mitral valve; Papillary muscles; Surgery.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Chordae Tendineae
  • Mitral Valve Insufficiency* / diagnostic imaging
  • Mitral Valve Insufficiency* / surgery
  • Mitral Valve* / diagnostic imaging
  • Mitral Valve* / surgery
  • Papillary Muscles
  • Swine
  • Vacuum