Aim: To determine inter-reader analysis and diagnostic performance on digitally reconstructed virtual flexed, abducted, supinated (FABS) imaging from three-dimensional (3D) isotropic elbow magnetic resonance imaging (MRI).
Materials and methods: Six musculoskeletal radiologists independently evaluated elbow MRI images with virtual FABS reconstructions, blinded to clinical findings and final diagnoses. Each radiologist recorded a binary result as to whether the tendon was intact and if both heads were visible, along with a categorical value to the type of tear and extent of retraction in centimetres where applicable. Kappa and interclass correlation (ICC) were reported with 95% confidence intervals. Areas under the receiver operating curve (AUC) were reported.
Results: FABS reconstructions were obtained successfully in all 48 cases. With respect to tendon intactness, visibility of both heads, and type of tear, the Kappa values were 0.66 (0.53-0.78), 0.24 (0.12-0.37), and 0.55 (0.43-0.66), respectively. For the extent of retraction, the ICC was 0.85 (0.79-0.91) when including the tendons with and without retraction and 0.78 (0.61-0.91) when only including tendons with retraction. For tear versus no tear, AUC values were 0.82 (0.74-0.89) to 0.96 (0.91-1.01).
Conclusion: Digital reconstruction of FABS positioning is feasible and allows good assessment of individual tendon head tears and retraction with high diagnostic performance.