Technical and clinical considerations of a physical liver phantom for CT radiomics analysis

J Appl Clin Med Phys. 2024 Apr;25(4):e14309. doi: 10.1002/acm2.14309. Epub 2024 Feb 22.

Abstract

Objective: This study identifies key characteristics to help build a physical liver computed tomography (CT) phantom for radiomics harmonization; particularly, the higher-order texture metrics.

Materials and methods: CT scans of a radiomics phantom comprising of 18 novel 3D printed inserts with varying size, shape, and material combinations were acquired on a 64-slice CT scanner (Brilliance 64, Philips Healthcare). The images were acquired at 120 kV, 250 mAs, CTDIvol of 16.36 mGy, 2 mm slice thickness, and iterative noise-reduction reconstruction (iDose, Philips Healthcare, Andover, MA). Radiomics analysis was performed using the Cancer Imaging Phenomics Toolkit (CaPTk), following automated segmentation of 3D regions of interest (ROI) of the 18 inserts. The findings were compared to three additional ROI obtained of an anthropomorphic liver phantom, a patient liver CT scan, and a water phantom, at comparable imaging settings. Percentage difference in radiomic metrics values between phantom and tissue was used to assess the biological equivalency and <10% was used to claim equivalent.

Results: The HU for all 18 ROI from the phantom ranged from -30 to 120 which is within clinically observed HU range of the liver, showing that our phantom material (T3-6B) is representative of biological CT tissue densities (liver) with >50% radiomic features having <10% difference from liver tissue. Based on the assessment of the Neighborhood Gray Tone Difference Matrix (NGTDM) metrics it is evident that the water phantom ROI show extreme values compared to the ROIs from the phantom. This result may further reinforce the difference between a structureless quantity such as water HU values and tissue HU values found in liver.

Conclusion: The 3-D printed patterns of the constructed radiomics phantom cover a wide span of liver tissue textures seen in CT images. Using our results, texture metrics can be selectively harmonized to establish clinically relevant and reliable radiomics panels.

Keywords: computed tomography; imaging; phantom study; radiomics; radiomics phantom.

MeSH terms

  • Humans
  • Image Processing, Computer-Assisted / methods
  • Liver / diagnostic imaging
  • Phantoms, Imaging
  • Radiomics*
  • Tomography Scanners, X-Ray Computed
  • Tomography, X-Ray Computed* / methods
  • Water

Substances

  • Water