Performance testing of the environmental TLD system for the Three Mile Island Nuclear Station

Health Phys. 1984 May;46(5):1013-20. doi: 10.1097/00004032-198405000-00002.

Abstract

Panasonic UD-801 thermoluminescent dosimeters ( TLDs ) containing two calcium sulfate phosphors were tested under Performance Specification 3.1 established by the American National Standard Institute ( ANSI75 ) and in the U.S. Nuclear Regulatory Commission's Regulatory Guide 4.13 ( NRC77 ). The specific qualifying tests included TLD uniformity, reproducibility, energy dependence and directional dependence. The overall measurement uncertainties and associated confidence levels are within the prescribed guidelines defined in the qualifying requirements for environmental TLDs .

MeSH terms

  • Accidents*
  • Nuclear Reactors*
  • Pennsylvania
  • Radioactive Pollutants / analysis*
  • Thermoluminescent Dosimetry / standards*