High-performance liquid chromatographic-ultraviolet assay for the simultaneous quantitation of BMS-181101 and its putative hydroxy metabolites in rat and monkey plasma

Biomed Chromatogr. 1996 May-Jun;10(3):135-8. doi: 10.1002/(SICI)1099-0801(199605)10:3<135::AID-BMC575>3.0.CO;2-B.

Abstract

A specific, accurate, precise, and reproducible High-performance liquid chromatographic-Ultraviolet (HPLC-UV) method was developed for the simultaneous quantitation of BMS-181101 (I), a new antidepressant, and its putative metabolites, 6'-hydroxy (II) and 7'-hydroxy (III) of BMS-181101 in rat and monkey plasma. The assay procedure involved solid-phase extraction of the three analytes and the internal standard (IS; BMY-42568) on 1 mL Bond Elut CN cartridge using an automated solid phase extraction controller (ASPEC) system. The final elution of the analytes was performed using 0.25% triethylamine in methanol. The eluate mixture was evaporated to dryness, the residue was reconstituted in the mobile phase and injected onto a Zorbax Phenyl column (4.6 x 250 mm; 5 microns) at a flow-rate of 1.2 mL/min. The mobile phase consisted of 20% acetonitrile, 10% methanol, 69% water and 1% 1.0 M ammonium phosphate and 1.0 M tetramethylammonium hydroxide mixture adjusted to pH 3 by phosphoric acid. An ultraviolet absorbance detector set at 287 nm was used to detect the analytes. The nominal retention times were 5, 8, 15, and 18 min for II, III, I, and IS, respectively. The standard curves for the three analytes were linear in the concentration range of 50-1000 ng/mL. The lower limit of quantitation was 50 ng/mL for each analyte. The analyses of quality control (QC) samples indicated that the nominal values could be predicted with an accuracy of (+/-) 10.5% for all three analytes in rat and monkey plasma. The precision values of the QC samples for all three analytes were within 12.7% RSD for rat and monkey plasma. All three analytes and the IS were stable in the autosampler for at least 38 h; freeze/thaw stability of the 3 analytes was established for three cycles. Stability of BMS-181101 was established for one month at -20 degrees C. The application of the assay to a pharmacokinetic study in monkey is described.

MeSH terms

  • Animals
  • Antidepressive Agents / blood*
  • Antidepressive Agents / metabolism*
  • Calibration
  • Chromatography, High Pressure Liquid / methods*
  • Drug Stability
  • Haplorhini
  • Hydrogen-Ion Concentration
  • Hydrolysis
  • Piperazines / blood*
  • Piperazines / metabolism*
  • Pyrimidines / blood*
  • Pyrimidines / metabolism*
  • Rats
  • Reference Standards
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrophotometry, Ultraviolet / methods
  • Temperature

Substances

  • Antidepressive Agents
  • BMS 181101
  • Piperazines
  • Pyrimidines