Estimation of total body and extracellular water in post-coronary artery bypass graft surgical patients using single and multiple frequency bioimpedance

Crit Care Med. 1996 Nov;24(11):1824-8. doi: 10.1097/00003246-199611000-00011.

Abstract

Objective: To assess the value of bioimpedance as a clinical tool by determining the accuracy and bias of single and multiple frequency bioimpedance estimates of total body and extracellular water in comparison with values established by criterion reference techniques.

Design: Controlled, prospective, single-blind investigation.

Setting: Private, not-for-profit, university-affiliated, acute care hospital.

Patients: Eight male, post-elective coronary artery bypass graft surgical patients.

Interventions: Within 6 hrs after surgery, estimates of total body and extracellular water volumes were determined using single and multiple frequency bioimpedance techniques. These estimates were then compared with the gold standard volumes measured by deuterium oxide and bromine dilutional space determination, respectively.

Measurements and main results: The mean multiple frequency bioimpedance estimate of total body water of 47.7 +/- 9.4 L was statistically different from the single frequency bioimpedance and deuterium values of 52.5 +/- 9.4 (p < .006) and 53.3 +/- 11.6 L (p < .002), respectively. In comparison, the mean multiple and single frequency bioimpedance estimates of extracellular water, 26.3 +/- 5.4 and 29.2 +/- 5.4 L, respectively, were not statistically different from the bromine value of 27.5 +/- 6.9 L. In addition, the mean errors for multiple and single frequency bioimpedance determinations of extracellular water, -1.2 +/- 2.0 and 1.7 +/- 2.7 L, respectively, were statistically different (p = .001).

Conclusions: In male, post-elective coronary artery bypass graft surgical patients, single frequency bioimpedance was a more accurate and less biased predictor of total body water than multiple frequency bioimpedance. The accuracy and bias of multiple frequency bioimpedance was superior to single frequency bioimpedance for the prediction of extracellular water. Whether this observation remains true for other populations of critically ill patients remains to be investigated.

Publication types

  • Clinical Trial
  • Comparative Study
  • Controlled Clinical Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Body Composition*
  • Body Water*
  • Coronary Artery Bypass*
  • Electric Impedance*
  • Extracellular Space*
  • Humans
  • Male
  • Middle Aged
  • Postoperative Period
  • Prospective Studies
  • Single-Blind Method