Tissue fixation security in transosseous rotator cuff repairs: a mechanical comparison of simple versus mattress sutures

Arthroscopy. 1996 Dec;12(6):704-8. doi: 10.1016/s0749-8063(96)90174-7.

Abstract

The primary purpose of this investigation was to compare tissue fixation security by simple sutures versus mattress sutures in transosseous rotator cuff repair. These two repair techniques were each performed in 17 human cadaver shoulders, with two bone tunnels being used for the repair by two simple sutures and two other bone tunnels being used for the repair by one mattress suture. The repairs were loaded to failure in a servohydraulic materials test system. Rotator cuff repair by simple sutures was found to be significantly stronger than repair by mattress sutures (P = .0007). The average ultimate load to failure for the simple suture construct (189.62 N) was 39.72% greater than that for the mattress suture construct (135.71 N). Most of the failures occurred by suture breakage at the knot. Load-sharing by multiple suture tails and multiple knots in the simple suture configuration likely contributed to its superior strength characteristics compared with the mattress suture configuration.

Publication types

  • Comparative Study

MeSH terms

  • Biomechanical Phenomena
  • Cadaver
  • Humans
  • Rotator Cuff / physiology
  • Rotator Cuff / surgery*
  • Suture Techniques*
  • Sutures*
  • Tensile Strength