Report of the ad hoc committee of the AAPM radiation therapy committee on 125I sealed source dosimetry

Int J Radiat Oncol Biol Phys. 1998 Feb 1;40(3):697-702. doi: 10.1016/s0360-3016(97)00767-0.

Abstract

Purpose: Two developments in 125I-sealed source dosimetry have necessitated swift and accurate implementation of TG43 dosimetry in clinic: (a) the dosimetry constants of 125I endorsed by the AAPM Task Group 43 Report result in calculated dose rate that deviates by as much as 15% from currently accepted dose-rate distributions, and (b) The National Institute of Standards and Technology (NIST) has proposed modifying the 125I air-kerma strength standard by approximately 10%.

Methods and materials: The ad hoc committee of AAPM Radiation Therapy Committee describes specific procedures to implement these two developments without causing confusion and mistakes.

Conclusions: Confusion and mistakes may be avoided when the following two general steps are taken: 1) STEP I, TG-43 implementation, and 2) STEP II, new air-kerma strength standard implementation when available from NIST.

Publication types

  • Guideline
  • Practice Guideline

MeSH terms

  • Brachytherapy / instrumentation
  • Brachytherapy / standards*
  • Humans
  • Iodine Radioisotopes / therapeutic use*
  • Radiation Oncology / standards*
  • Radiotherapy Dosage / standards
  • Reference Standards